Defibrillators
A defibrillator is a medical device used in the defibrillation of the heart. It consists of a central unit and a set of two electrodes. The central unit provides a source of power and control. The two electrodes are placed directly on or in the patient. The device is designed to deliver an electric shock to the patient, in an effort to stop ventricular fibrillation.
Contents 1 Types of Defibrillators 1.1 Internal Defibrillators 1.2 External Defibrillators 1.2.1 Biphasic Defibrillation 1.2.2 Automated External Defibrillators 2 Electrodes 2.1 Design 2.2 Placement 2.3 Inventor
Types of Defibrillators
Internal Defibrillators The device may be implanted directly in the user of the device. If so, it is known as an implantable cardioverter-defibrillator or (much less frequently) an internal cardiac defibrillator (ICD). This type of defibrillator is designed to provide immediate defibrillation to high-risk patients. By actively monitoring the pulse rate, rhythm, and waveform, and by comparing atrial and ventricular activity, in ICD can detect ventricular fibrillation, and immediately initiate defibrillation.
External Defibrillators External defibrillators are typically used in hospitals or ambulances, but are increasingly common outside the medical realm, as automated external defibrillators become safer and cheaper. There are a variety of technologies and form factors in use for external defibrillators, and recent progress in cardiac research has led to substantial improvements in the underlying technology.
Biphasic Defibrillation Until recently, external defibrillators relied on monophasic shock waves. Electrical pulses are sent rapidly from one electrode to the other, only in one direction. Biphasic defibrillation, however, alternates the direction of the pulses, completing one cycle in approximately 10 milliseconds. Biphasic defibrillation was originally developed and used for implantable cardioverter-defibrillators. When applied to external defibrillators, biphasic defibrillation significantly decreases the energy level necessary for successful defibrillation. This, in turn, decreases risk of burns and myocardial damage. However, there is limited evidence to suggest that biphasic defibrillation is superior to monophasic defibrillators, although the small capacitor size required for the defibrillator can result in significant cost and size savings-- essential for the proliferation of Automated External Defibrillators.
Automated External Defibrillators An Automated External Defibrillator (AED) is a self-contained defibrillator device designed for portability and ease of use. AEDs are generally shaped like a briefcase, so that they may be carried easily by a handle. An AED contains a battery, a control computer, and electrodes. Upon placing the electrodes on the patient, the control computer in an advanced system will assess the patient, determining the type of rhythm or arrhythmia present. It will then set appropriate power levels and signal that a shock is needed. If the patient does not require defibrillation, units will not allow a shock to be administered. A button must still be depressed manually to trigger the shock, as the operator beforehand must be certain that no-one is touching the patient at that time.
Current AED devices are designed for emergency medical technicians, home users, public safety officers and other people with minimal medical knowledge. AEDs are available for $1000 for a basic model to several thousand dollars for a more fully-featured or durable model.
These devices are commonly found in large gathering places, such as airports, casinos, sports stadiums, and college campuses.
Electrodes The electrode is a key part of any defibrillation system. The proper selection and placement of electrodes can determine the effectiveness of the procedure.
Design The most well-known type of electrode is the traditional metal paddle with an insulated handle. This type must be held in place on the patient's skin while a shock or a series of shocks is delivered. Before the paddle is used, a gel must be applied to the patient's skin, in order to ensure a good connection and to minimize electrical impedance.
Another type of resuscitation electrode is designed as an adhesive pad. When a patient has been admitted due to heart problems, and the physician or nurse has determined that he or she is at risk of arrhythmia, they may apply adhesive electrodes to the patient in anticipation of any problems that may arise. These electrodes are left connected to a defibrillator. If defibrillation is required, the machine is charged, and the shock is delivered, without any need to apply any gel or to retrieve and place any paddles.
Both solid- and wet-gel adhesive electrodes are available. Solid-gel electrodes are more convenient, because there is no need to clean the patient's skin after removing the electrodes. However, the use of solid-gel electrodes presents a higher risk of burns during defibrillation, since wet-gel electrodes more evenly conduct electricity into the body.
Adhesive electrodes are designed to be used not only for defibrillation, but also for non-invasive pacing and electrical cardioversion.
While the paddles on a Monitor/Defibrillator may be quicker than using the patches, adhesive patches are superior due to their ability to provide appropriate EKG tracing without the artifact visible from human interference with the paddles. Many monitor defibrilators provide three, five or 12-lead EKG monitoring to compensate for this downfall of the paddles. Another inconvenience of the paddles is the requirement of around 25lbs of pressure to be applied while defibrillating.
Placement Resuscitation electrodes are placed according to one of two schemes. The anterior-posterior scheme is the preferred scheme for long-term electrode placement. One electrode is placed over the left precordium (the lower part of the chest, in front of the heart). The other electrode is placed on the back, behind the heart in the region between the scapula. This placement is preferred because it is best for non-invasive pacing.
The anterior-apex scheme can be used when the anterior-posterior scheme is inconvenient or unnecessary. In this scheme, the anterior electrode is placed on the right, below the clavicle. The apex electrode is applied to the left side of the patient, just below and to the left of the pectoral muscle. This scheme works well for defibrillation and cardioversion, as well as for monitoring an ECG.
Inventor The first case of a human life saved by defibrillation was reported by Beck et al in 1947. Claude Beck successfully revived a patient in an operating room using an open-chest electric defibrillation device.
The closed-chest defibrillator device was pioneered by Dr V. Eskin with assistance by A. Klimov in the city of Frunze, USSR in mid 1950s. Successors of this device continue to be used to this day.
Another advance was the development of a mobile defibrillator in 1966. The device was suitable for installation in ambulance vehicles and was developed by cardiologist Frank Pantridge in Belfast, Ireland. This is one of the innovations that led to modern EMS and, in the late 1990s, the mobilization of advanced cardiac life support with paramedics.
|